The transformer calibration set is an important equipment for the error verification of current transformers. This device consists of a voltage regulator, a transformer load box, a standard transformer, and a transformer calibration instrument, serving as the basis for the error verification work of transformers.
The following equipment must be provided according to actual requirements: voltage regulators, load boxes for current transformers, standard current transformers, and current transformer calibration instruments.
The Transformer Calibration Device is a next-generation automatic
testing platform for voltage regulators developed by our company.
Utilizing cutting-edge virtual instrument design principles, users
can perform fully automated testing of current transformers and
voltage transformers through computer interfaces using mouse and
keyboard operations. The system enables comprehensive analysis,
data processing, and database management of test results, with
customizable printing of test reports and certification documents
tailored to user requirements. This equipment serves as an ideal
solution for transformer calibration in power systems and related
industries. Compared to domestic calibration platforms, its key
features include: 1. True virtual instrument architecture with
fully automated control – users simply input parameters such as
transformer type, measurement range, accuracy class, capacity, and
power factor via computer interfaces to complete the entire
workflow from measurement to certificate generation.
2. Powered by a variable frequency source, it achieves precise,
rapid, and stable positioning, completely resolving the 1%
positioning challenge for S-class current transformers.
3. The voltage and current load box automatically switches gear
positions through programmed control, eliminating the need for
manual calculations or gear shifts to prevent incorrect gear
selection.
4. Equipped with comprehensive alarm and overcurrent protection
functions.
5. The included transformer calibration instrument features a
built-in touch screen, greatly facilitating debugging,
verification, and field operations.
Key Technical Specifications of the Device
1.1 Main Technical Specifications for Programmable Source
Calibration Bench:
1.1.1 To achieve stable and rapid 0-250V output voltage, the
calibration bench must utilize a fully automated programmable
source equipped with integrated display and protection features.
Power input for the transformer: Voltage: AC 220V±10%;
Frequency: 50Hz±0.5 Hz.
1.1.2 Operating environment: Temperature: -10℃ to 40℃; Humidity:
<80% (at 25℃).
1.1.3 Output power: 10kVA.
1.1.4 Accuracy of percentage meter: 1%, Display range: 1% to 199%.
1.1.5 Secondary measurement range: Voltage: 100V,100V/; Current:
1.5 A
1.1.6Automatic stability point:
Voltage (%): 20,50,80,100,120
Current (%): 1.5, 20,100,120
1.1.7Overcurrent protection setting: Automatic
1.1.8 Minimum adjustment: <0.04%
1.1.9 Weight: <120 kg
1.2 Transformer calibration instrument parameters
1.2.1 Measurement range:
Same Phase component (%): 0.0001–200.0 Resolution: 0.0001
Orthogonal component (°): 0.001–999.9 Resolution: 0.001
Impedance (Ω): 0.0001–10.0 Resolution: 0.0001
Conductance (S·s): 0.0001–10.0 Resolution: 0.0001
1.2.2 Fundamental error:
Same Phase component: (X = (X × 2% + Y × 2%)(D)
Orthogonal component: (Y = (X × 2% + Y × 2%)(D)
"X" and "Y" – Instrument display values
"D" – Instrument quantification error 0.5 class
1.2.3 Operating range:
Current: (1%–149% In) (In = 5A) (5%–149% In) (In = 1A)
Voltage: (5%–149% Un) (Un = 100V) 100V/) (5%–199%) Un (Un=100V/)
1.2.4 Operating load: Current: To For Tx<0.12 (cosφ=1 Voltage: a For x<0.25VA (100V)
1.2.5 Polarization error indication When the rated operating
current (voltage) deviation exceeds 5% and the error exceeds 180%,
polarity indication must be provided.
1.2.6 Transformer ratio error indication shall be triggered when
the error exceeds 5% of the rated working current (voltage), with
the error range between 30% and 180%.
1.2.7 Insulation and Voltage Withstand Test Specifications:
Terminal Tx is connected to terminal (), while terminals K and D
are isolated from terminal (). The power socket withstands 1.5kV
voltage for 1 minute.
1.3 Automatic Current-Voltage Transformer Load Box Parameters: Both current and voltage ranges can be automatically switched.
1.3.1 Rated Capacity:
Voltage: 1.25VA~158.75VA (COSΦ=0.8, COSΦ=1.0);
Current: 2.5VA~60VA (COSΦ=0.8, 2.5VA~5VA COSΦ=1.0).
1.3.2 Measurement Range:
Voltage: 20%~120%; Current: 1%~120%.
1.3.3 Secondary Voltage: 100V,110V,100/√3V,110V/√3V;
Secondary Current: 5A,1A.
1.3.4 Power Factor: 0.8.
1.3.5 Accuracy: ±3%.
5.4 High-Current Precision Standard Current Transformer:
The standard current transformer equipped with a current booster serves dual functions as a measurement device and reference standard transformer. Technical specifications:
Primary current: 5A~5000A;
Secondary current: 5A,1A;
Frequency: 50Hz;
Accuracy class: Class 0.05S--0.002S;
Booster voltage: 250V;
Booster capacity: 20kVA;
Rated load: 5VA. 1VA
lower limit load: 2.5VA
Power factor: 1.0
Rated voltage: 500V
5.4:Standard voltage transformer
Key features:
The voltage transformer is a high-voltage, high-accuracy device
capable of calibration for transformers with accuracy below 0.1
class. It can handle a secondary load of 0.25VA at 100V, fully
meeting the calibration requirements for voltage transformers.
Technical specifications: Rated voltage ratio
10kV/100V--220kV/100V,
rated secondary capacity 0-0.20VA,
power factor (Cosφ) = 1.0,
accuracy class 0.05--0.002,
rated frequency 50Hz.
|