Sign In | Join Free | My infospaceinc.com
infospaceinc.com
Products
Search by Category
Home > Measuring & Analysing Instrument Design Services >

laboratory Crystallinity Thermal Analyzer AC220V/50Hz Or Customized

Categories Pharmaceutical Testing Instruments
Brand Name: Shandong Shengtai Instrument
Model Number: ST146
Place of Origin: CHINA
MOQ: 1
Payment Terms: T/T
Supply Ability: 10000 sets/year
Delivery Time: 7-15 work days
Packaging Details: wooden case
Temperature range: room temperature~600℃
Temperature sensitivity: 0.001℃
Temperature fluctuation: ± 0.01℃
Temperature repeatability: ± 0.1℃
Heating rate: 0.1~100 ℃/min
Data scanning: heating scan, cooling scan, constant temperature scan
Temperature control method: PID precise control
DSC range: 0~± 600mW
DSC resolution: 0.01uW
DSC sensitivity: 0.001mW
Working power supply: AC220V/50Hz or customized
Atmosphere control gas: nitrogen,oxygen(instrument automatically switches)
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
Submit Buying Request
  • Product Details
  • Company Profile

laboratory Crystallinity Thermal Analyzer AC220V/50Hz Or Customized

ST146 crystallinity thermal analyzer

ST146 crystallinity thermal analyzer is designed and manufactured based on the new version of the General Principles of the 2020 Chinese Pharmacopoeia 0981 for crystallinity detection and the 0661 differential scanning calorimetry method. The crystallinity thermal analyzer adopts a touch screen design and can perform tests such as glass transition temperature, phase change, melting and enthalpy values, product stability, curing, specific heat, and oxidation induction period

Technical parameters and characteristics

Temperature rangeroom temperature~600℃
Temperature sensitivity0.001℃
Temperature fluctuation± 0.01℃
Temperature repeatability± 0.1℃
Heating rate0.1~100 ℃/min
Data scanningheating scan, cooling scan, constant temperature scan
Temperature control methodPID precise control
DSC range0~± 600mW
DSC resolution0.01uW
DSC sensitivity0.001mW
Working power supplyAC220V/50Hz or customized
Atmosphere control gasnitrogen,oxygen(instrument automatically switches)
Gas flow rate0-300mL/min
Display method24 bit color, 7-inch LCD touch screen display
Parameter standardequipped with standard substance (tin), users can calibrate the temperature themselves

● Temperature range: room temperature~600 ℃

● Temperature sensitivity: 0.001 ℃

● Temperature fluctuation: ± 0.01 ℃

● Temperature repeatability: ± 0.1 ℃

● Heating rate: 0.1~100 ℃/min

● Data scanning: heating scan, cooling scan, constant temperature scan

● Temperature control method: PID precise control (fully automatic program control); The lower computer interface can set and define PID parameter values to meet different environments and power heating needs; Customizable initial heating power;

● DSC range: 0~± 600mW

● DSC resolution: 0.01uW

● DSC sensitivity: 0.001mW

● Working power supply: AC220V/50Hz or customized

● Atmosphere control gas: nitrogen, oxygen (instrument automatically switches)

● Gas flow rate: 0-300mL/min;

● Display method: 24 bit color, 7-inch LCD touch screen display

● Parameter standard: equipped with standard substance (tin), users can calibrate the temperature themselves

● Software: With specific heat capacity testing function, including specific heat capacity testing standard samples; The software can simultaneously set 6 experimental parameters;

● The software comes with automatic analysis function, data can be generated in Excel format, and experiments can be saved as PDF reports

● An industrial grade 7-inch touch screen with rich display information.

● New metal furnace body structure, better baseline and higher accuracy. Heating adopts an indirect conduction method, which has high uniformity and stability, reduces pulse radiation, and is superior to traditional heating modes.

● USB communication interface, strong universality, reliable and uninterrupted communication, supporting self recovery connection function.

★ Data extension: Account layering and audit tracking function (optional)

Note: Gas self provided

Packing list

NONameSpecificationsQty
1initial adhesion testerHost1
2USB flash drive1
3Aluminum crucible
4Pure tin particles
5Data cable
6Computer
7Power line1
8Instructionpiece1
9warranty cardPiece1
1

Shandong Shengtai Instrument Co., Ltd. provides the following quality assurance for the instruments sold to you:

----The provided instrument materials are brand new, comply with national quality standards, and have the manufacturer's certificate of conformity;

----The materials and main components provided meet the technical requirements specified in the technical documentation;

----The warranty period for the entire equipment is one year (excluding normal wear and tear of vulnerable parts).

----We are responsible for free maintenance of any instrument quality issues that occur during the warranty period. Due to the user's responsibility for causing equipment malfunction, we are responsible for maintenance and reasonable charges.

----Lifetime discounted supply of components for equipment, and lifelong maintenance and repair of the entire machine.

----After the expiration of the warranty period, if the user needs maintenance and technical services, we will only charge a cost fee.

Buy laboratory Crystallinity Thermal Analyzer AC220V/50Hz Or Customized at wholesale prices
Send your message to this supplier
 
*From:
*To: Shandong Shengtai instrument co.,ltd
*Subject:
*Message:
Characters Remaining: (0/3000)
 
Inquiry Cart 0