Sign In | Join Free | My infospaceinc.com
infospaceinc.com
Products
Search by Category
Home > Laser Equipment Parts >

Maintaining Airflow and Pressure Control Laminar Flow Hood

Categories Laminar Flow Hood
Brand Name: GCC
Model Number: GCC-FG-1000
Certification: ISO/GMP/CE
Place of Origin: Guangdong,China
MOQ: 1
Price: 1800$
Payment Terms: T/T
Supply Ability: 5000Sets/Month
Delivery Time: 14Days
Packaging Details: Wooden Box
Product name: Laminar Flow Hood
Material: Stainless Steel
Function: Contamination Control
Usage: Particle Filtration
Feature: Durability and Longevity
Advantage: Maintaining Airflow and Pressure Control
Applicable Industries: Manufacturing Plant
Video outgoing-inspection: Provided
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
Submit Buying Request
  • Product Details
  • Company Profile

Maintaining Airflow and Pressure Control Laminar Flow Hood

Product Name: Laminar Flow Hood

Dimensions: 18009002600 (mm)

Cleanliness Level: Static Class 100 @ 0.3um (US Federal Standard)

Cabinet Material: SUS304 Stainless Steel

Average Face Velocity: 0.25~0.55m/s (adjustable in three levels), recommended 0.35m/s

HEPA Filters: 2 pcs, 78969369 (mm), efficiency 99.995%@≥0.3μm

Medium Efficiency Filters: 4 pcs, 75025021, efficiency 45%@≥0.5μm

Pre-filters: 4 pcs, 75025017, efficiency 40%@≥2.0μm

Noise Level: ≤65db

Power Supply: 220VAC, 50HZ, 550W (excluding power consumption of sockets)

Note: Pre-filters should be replaced every 1-3 months, medium efficiency filters every 3-6 months, and HEPA filters generally every year.

Replacement Guidelines:

  • HEPA filters must be replaced when the differential pressure reaches 400Pa.
  • Pre-filters must be replaced or cleaned when the differential pressure reaches 50Pa.
  • Medium efficiency filters must be replaced or cleaned when the differential pressure reaches 80Pa.

Key Features:

NoItemDetail
1Cleanliness LevelStatic Class 100 @ 0.3um (US Federal Standard)
2Cabinet MaterialSUS304 Stainless Steel
3Average Face Velocity0.25~0.55m/s (adjustable in three levels), recommended 0.35m/s
4HEPA Filters2 pcs, 78969369 (mm), efficiency 99.995%@≥0.3μm
5Medium Efficiency Filters4 pcs, 75025021, efficiency 45%@≥0.5μm
6Pre-filters4 pcs, 75025017, efficiency 40%@≥2.0μm
7Size(W*D*H)MM1300*1200*600/1500*1200*600/2400*1200*600/2280*1460*500/1530*1060*500/1160*775*500
8NotePre-filters should be replaced every 1-3 months, medium efficiency filters every 3-6 months, and HEPA filters generally every year.

FAQ:

Q1:Are you a manufacturer or a trader? Where is your factory?
A1: We are a professional Purification Equipment manufacturer with a professional team of designers and engineers. Our factory is located in Guangzhou, China.

Q2:What is your delivery time?
A2: It will take 10-15 days after receiving your prepayment. The specific delivery time depends on your order quantity.