Product Overview
The cold bending testing machine is one of the climate simulation
chambers, suitable for testing the adaptability of various products
or materials, household appliances, instruments, meters, components
and other electrical and electronic products under high and low
temperature environments. It can be used for testing in various
scientific research institutions and factory and mine central
laboratories. This testing machine is suitable for cold bending
tests of wires. First, set the temperature of the test chamber.
When the temperature reaches the set value, take a sample of a
certain length and put it in. After the test time, immediately wind
it around a circular rod of the specified diameter at a uniform
speed, then take out the sample and check if there are any cracks
or fractures.
Product images
Product features
The unique balanced temperature regulation method enables the
equipment to achieve high-precision and highly stable temperature
control.
The studio is made of high-quality stainless steel plates, and the
sample shelves are also made of stainless steel, which is
corrosion-resistant and easy to clean.
3. The temperature controller of the equipment features high
precision and stability, PID self-tuning and LED intuitive display,
ensuring precise control of the equipment.
The equipment is equipped with over-temperature protection and
timing functions. When the timing ends or an alarm is triggered,
the power supply will be automatically cut off to stop the
operation of the equipment, ensuring the safety of the equipment
and personnel.
The sealing strips of the equipment are made of silicone material,
which features good toughness and is not prone to deformation or
stickiness in high and low temperature environments.
The box body is made of stainless steel, with uniform color and a
beautiful and elegant appearance.
Product features
Execution standards and main technical parameters
This testing machine complies with the test standards of GB2951.4,
UL62, IEC540 and VED. The specimens were wound in a specific
low-temperature environment according to different rod diameters to
check whether cracks occurred on their surfaces.
Temperature range: -40℃ to RT
The working dimensions are 50×50×40cm
The diameters of the round rods are: 4 each of ¢6.5mm, ¢12.7mm,
¢19.0mm, and ¢25.4mm.
Product usage conditions (optional)
<s:1> Temperature: 15-35℃;
Relative humidity: not more than 85%R.H;
Atmospheric pressure :86-106Mpa;
Dielectric voltage: AC 220V
Reactive power: 3KW
There is no strong vibration around. '
There is no direct sunlight or direct radiation from other heat
sources.
There should be no strong air flow around. When the surrounding air
is forced to flow, the air flow should not directly blow onto the
box.
There is no strong electromagnetic field influence around.
There should be no high-concentration dust or corrosive substances
around.
To ensure the normal operation of the equipment and the convenience
of operation, in addition to keeping the equipment placed
horizontally, a certain space should be reserved between the
equipment and the wall or objects.
Safety Precautions
To protect the equipment and ensure the safety of the test, please
install an external protective mechanism.
The equipment should be installed horizontally and it is strictly
prohibited to place it at an Angle.
Non-professionals are not allowed to disassemble or repair the
equipment system.
During the test, do not open the box cover unless it is absolutely
necessary, otherwise it may cause personal injury and malfunction
of the system.
The equipment should be reliably grounded to ensure safety and the
effective control of the protection control system.
If a heating sample is placed inside the device, please use an
external power supply for the sample. Do not directly use the
device's power supply.
Tests on explosive, flammable and highly corrosive substances are
strictly prohibited.
The power circuit breaker of the equipment, the over-temperature
protection system, the provision of equipment test pieces, and the
safety protection of the operators should be inspected regularly.
You can operate this equipment only after reading the manual
carefully.
When the machine is operating below 0℃, the box door should not be
opened. Otherwise, frost and water will form on the evaporator and
other parts inside the box, reducing the efficiency of the machine.
After the low-temperature operation is completed, please set this
machine to 60℃ and conduct a 30-minute drying operation.
This machine is equipped with a timer switch. When the temperature
inside the box reaches the test temperature, if timed operation is
required, please turn it on. After the set time is reached, the
machine will automatically cut off the power supply and send out a
prompt signal.
During the youdaoplaceholder3 test, if the box door is opened, it
may cause misunderstanding of the system PID and trigger the
over-temperature protection. Please avoid this.
After the cold machine switch is turned on, the equipment enters a
delayed protection state. Do not forcibly start it in this state.
Do not change the set parameter values during the operation of the
instrument; otherwise, it will be difficult to ensure the control
accuracy of the equipment.
This equipment is equipped with an observation window and a
lighting device for observing the test conditions inside the
laboratory. This lighting device is only used for observation.
|